Quantum-data 801GC Manual de usuario Pagina 228

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 356
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 227
6-102 Chapter 6: Programming
Model 801GC, 801GF & 801GX¥Rev. A
volts. The <sync 400> factor adjusts the analog sync output level
when ASSS is at 0.400 volts. The <sync 40> factor adjusts the
analog sync output level when ASSS is at 0.040 volts. The <NTSC
714> factor adjusts the NTSC television output level. The <PAL
700> factor adjusts the PAL television output level. The <pclk
700> factor adjusts the pixel clock output (when it is enabled).
Information on the <psync> factors was not available when this
manual was written.
Other Required Cmds: The AVSC command matches the levels for the three analog
video channels. The ALLU command updates the signal
generating hardware to the new settings and redraws the test
image.
Example: CALF 2040 2045 2055 2050 2046 2048
// Set new factors
ALLU // Use new factors