Quantum-data 801GC Manual de usuario Pagina 290

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 356
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 289
6-164 Chapter 6: Programming
Model 801GC, 801GF & 801GX¥Rev. A
SIZE SIZE of bit field
Classification: Direct processor control
Command Syntax: SIZE <size>
Limits: <size>
-32 to -1 or 1 to 32 bits
Query Syntax: SIZE?
Query Response: <size>
Description: The SIZE command sets the field size (in base 10) used in
connection with the GETA, GETR, PUTA, and PUTR commands.
If a negative size is specified, then values given (or returned)
are sign extended to 32 bits. For example, if SIZE = -16 and a
16 bit field containing FFFF hex is fetched, then the value
FFFFFFFF hex is returned. The SIZE? query returns the current
setting of SIZE in base 10. SIZE is preset to 16 each time the
generator is powered on.
Note This command will normally be used only with custom
applications and command files created by Quantum Data.
Other Required Cmds: None
Example: Use only with code supplied by Quantum
Data!
SMOD Sequence operating MODe
Classification: Sequence parameter setting
Command Syntax: SMOD
Limits: <mode>
0 = Disable
1 = Enable manual step mode that stops at last step
2 = Enable manual step mode that wraps to first step
after last step
3 = Enable continuous auto stepping
Query Syntax: SMOD?
Query Response: <mode>
Description: The SMOD command sets the sequence mode. The SMOD? query
returns the current setting of SMOD.
Other Required Cmds: The SEQL command loads the sequence and SEQU starts to run
it.
Example: SEQL MY_SEQ // Load "MY_SEQ" dir in
// edit buffer
SEQU // Start running the sequence in
// the buffer
SMOD 3 // Set the sequence mode